APPENDIX C – MAINTENANCE
71
APPENDIX C – MAINTENANCE
In addition to correct use, the user needs to perform ordinary maintenance in order to guarantee safe, efficient operation
over the device's entire life.
INTRODUCTION
For better quality maintenance, supplement ordinary checks with regular periodic
examinations by the service department (see Appendix Z).
It is also fundamental to perform a periodic sterilizer validation, i.e., a check of the
thermodynamic parameters of the process, comparing them with the reference values
provided with suitably calibrated instruments. In this regard, see the paragraph, “Periodic
Sterilizer's Validation”, further below in this Appendix.
The ordinary maintenance described below consists in easy manual operations and preventive
interventions involving simple instruments.
WARNING
IN THE EVENT OF THE REPLACEMENT OF THE DEVICE'S COMPONENTS
OR PARTS, REQUEST AND/OR USE ORIGINAL REPLACEMENT PARTS
ONLY.
ORDINARY
MAINTENANCE
PROGRAM
Scheduled Maintenance
Messages
The table summarizes the maintenance required to keep the sterilizer operating at peak
efficiency. In the case of very intense use
, we recommend shortening maintenance
intervals:
DAILY
Clean the gasket on the porthole
Clean external surfaces
WEEKLY Clean the sterilization chamber and relative accessories
Disinfect external surfaces
MONTHLY Clean the internal (and external - if installed) distilled water tank
Safety valve maintenance
Clean (or replace) the drain filter
ANNUALLY
Validate sterilizer (see dedicated paragraph)
The steriliser periodically reminds the user about necessary "routine" maintenance operations
that must be carried out in order to ensure the proper operation of the device.
The alert notices are displayed on the screen in the following way when the pre-selected
sterilisation cycle is started:
(ALERT MESSAGES)
to confirm
to delay
Push the key to confirm the execution of the maintenance operation.
Press the key to postpone the operation.
In this case, the user is reminded with another message the next time the steriliser is used.