5. FIRST START-UP
18
STAND-BY MODE
After the initial test, the sterilizer goes to STAND-BY mode and the display shows:
Counterxxxxx/yyyyy
Stand‐ by HIGH
23.6 °C01/02/10
‐ 0.01 bar18:13:05
The upper line is the cycle counter for sterilizations performed, with the number of correctly
completed cycles on the left and the total number started on the right. The line below shows
the Stand-by status and the preheating mode (High-Low-Off). The two lower lines show the
temperature and pressure of the sterilization chamber on the left and current date and time on
the right.
NOTE
CYCLE BEGINS WITH THE START OF THE STERILIZATION CYCLE (FIRST VACUUM
PHASE
), EXCLUDING THE PREHEATING PHASE. A CYCLE ENDS AT THE END OF THE
PROGRAM
(SEE THE CHAPTER, “PROGRAM EXECUTION”).
TO SET THE DATE AND TIME AS WELL AS SELECT THE PREHEATING MODE, PRINT THE
DATA AND FILL THE TANK
, PLEASE REFER TO THE CHAPTER, “CONFIGURING THE
DEVICE
”.
At regular intervals, the first two lines on the display alternate with the modes set for printing
(ON/OFF) and filling (Manual/Automatic):
PrintON
Filling MANUAL
23.6 °C01/02/10
‐ 0.01 bar18:13:05
The icons in the lower part of the LCD screen remain off with the exception of the door status
and/or water level indicators, which light-up if the door is closed and/or the level in the filling
tank reaches its MIN or MAX values (or the MAX value in the drain tank).
During the first start-up, the MIN water level icon in the filing tank is normally on.
The device waits for the selection of the desired sterilization program (see the Chapter,
“Program Selection”).
DANGER
WHEN THE DOOR IS OPEN IN STAND-BY MODE, A BEEP INDICATES
THAT THE SURFACES INSIDE THE DEVICE ARE HOT. TO AVOID
BURNS, TAKE CARE NOT TO TOUCH THE STERILIZATION CHAMBER,
THE SUPPORTS PROVIDED OR THE INSIDE OF THE DOOR WITH YOUR
BARE HANDS.